DRY HEAT STERILIZER

SALIENT FEATURES :

  • Ambica Boiler & Fab. depyrogenation steriliser are capable of performing up to 280 ° C, during the entire cycle. We have achieved this, due to our unique chamber construction and filter suspension system, and our never ending work for further improvements.
  • High Temperature Hepa Filters High Temp HEPA filters are installed at inlet and exhaust side of the DHS. The HEPA-filters inside the chamber are equipped with full depth and height of chamber for suspension system solving the particle problem during heating and cooling.
  • Chamber The Ambica Boiler & Fab. depyrogenation steriliser is made of stainless steel 316L. The chamber is fully welded inside, no leakage to the insulation can occur and the chamber is very easy to clean. The water cooling element is a continuous welded unit without any connections inside the cabinet boundary. It is pressure tested and is expected to last the lifetime of the steriliser.
  • Door gasket As standard, Neelam Industries sterilizers are equipped with double door gaskets (FDA-approved). The purpose of double door gasket is to render maximum security, allowing no leakage between the chamber and the out­side.
  • Further more the double gasket minimizes the surface temperature of the sterilizer, and any undesired air flow in the sterile area.
  • Superior air control - Guarantee for complete sterilization Neelam Industries depyrogenation sterilisers operate in accordance with the forced convection principle with fast air circu­lation securing a rapid heat transmission to the load. Together with the shape and position of the electrical heating element, the calculated air distribution plates ensure a unique temperature distribution in the sterilization chamber. The temperature distribution is tested and fixed before delivery and does not require any adjustments in the life time of the steriliser, accordingly no down time for re-validation is needed. The uniform temperature in the chamber make it possible to fix the set-point temperature very close to the required minimum temperature for the sterilization.
  • Program Cycle The cycle of a ABF sterilizer is fully automatic.
  • Drying The cycle starts with the drying phase where the damper on the exhaust is open ensuring, that any moisture will be removed from the chamber.
  • Heating At a set-point temperature decided by the customer, the damper will close and the temperature will rise to the sterilizing set-point temperature. The damper will also stay closed during sterilization. Ambica Boiler & Fab. Sterlisers are equipped with a power-ful heating element to shorten the heating time
  • Sterilizing The sterilizing timer can be started by the load sensors. The sterilizing timer will only be running if all the load sensors are within an acceptable range from the sterilizing set-point.
  • Cooling If the cycle time is important, forced air cooling or water cooling is avail­able for rapid cooling in customized model.
Dry heat Sterilizer
 

Dry Heat Sterilizer

   
   

 

 
 
 